Setaflash Series 8 Activecool Flash Point Tester (corrosion resisting)
Part Number 82150-0
The Setaflash Series 8 'ActiveCool' is an automated closed cup flash point tester with a temperature range of 10 to 130°C (50 to 266°F) with no external cooling required. When connected to a suitable chilled water supply, the range is extended down to -20°C (-4°F).
There are two modes of operation; Flash / No Flash and Ramp. The ramp mode increases the sample temperature at a rate of 2°C/min and automatically dips the ignitor every 1°C up to 100°C (every 2°C above 100°C), until either a flash is detected or the end of test temperature is reached.
ADDITIONAL INFORMATION
The automatic flash detection system is tuned to ignore halo and other false flash effects, enhancing precision and accuracy. Barometric pressure is measured by an internal barometer, and the correction is automatically applied to the test result.
Peltier cell technology is used for heating and cooling, ensuring precise temperature control and rapid cooldown after a test is completed. The ignitor uses a new electrical hot wire (a gas flame option is available). The bright colour LCD display shows test parameters and the flash point result. The text and colour of the top line indicates instrument status. All functions are accessed via menus, navigated by the keypad. Up to 64 test results can be stored in memory, each tagged with a sample reference and operator name. Results can be downloaded to a PC or printer via the RS232C port (USB or LAN ports options available). The firmware is upgradeable for future developments.
SUPPLIED WITH: Ignitor, 2ml syringe, mains lead and instruction manual.
SPECIFICATIONS
| Temperature Range | Air: 10 to 130°C (50 to 266°F) Water Assisted: -20 to 130°C (-4 to 266°F) using 99300-2 |
| Sample Size | 2 or 4ml according to method |
| Ignitor | Electrical Hot Wire (optional Gas Flame) |
| Test duration, Rapid Equilibrium Mode | 1 minute below 100°C, 2 minutes above 100°C, FAME 1 minute or user defined 1 to 99 minutes |
| Test duration, Ramp Mode | Typically 7 minutes |
| Cool Down Time | 80 to 20°C (176 to 68°F) in 4 minutes |
| Sample Cup Material | Corrosion resisting steel insert, aluminium cup |
| Heating / Cooling Method | Peltier Cell |
| Computer Interface | RS232C (USB & LAN options) |
| Voltage | 100 to 250V, 50/60Hz (2x AA dry cell batteries required for internal clock) |
| Power | 300W (maximum) |
| Size (HxWxD) | 30 x 34 x 38cm |
| Weight | 8kg |